Operating Requirements
Electrical Ratings
Power Requirements
Power connector
The instrument has the following power requirements:
A single-phase power source with one current-carrying conductor at or near
earth-ground (the neutral conductor).
NOTE. Systems with both current-carrying conductors live with respect to ground
(such as
phase-to-phase in multiphase systems) are not recommended as power
sources.
The mains supply frequency must be 50 or 60 Hz.
The mains supply voltage must be in the range from 100 to 240 V
AC
.
WARNING. To reduce risk of fire and s hock, ensure the mains supply voltage
fluctuations do not exceed 10% of the operating voltage range.
Fus
es
Only the line conductor is fused for over-current protection. The fuse is internal
and not user replaceable. Do not attempt to replace the fuse. If you suspect the
fuse has blown, return the instrument to an authorized service center for repair.
Ba
tteries
The instrument does not contain any user-replaceable batteries.
Input Ratings
Table 2: Maximum input voltage
Input Rating
At front-panel BNC connector 300 V RMS, Installation Category II; derate above 4 M Hz to 6 V RMS at 200 MHz.
At the P6316 probe input, not at the
instrument input
±40 V peak
14 MSO2000B and DPO2000B Installation and Safety Instructions