Commands Listed in Alphabetical Order
CURSor:VBArs
:POSITIO N<x>
Sets or returns the horizontal position for the specied vertical bar cursor. The
cursor is specied by <x>, which can be 1 or 2. Values are with respect to trigger
position or t
he zero reference point for the designated waveform (if horizontal
units are not set to time). Use the CURSor:VBArs:UNIts command to specify
units.
Group
Cursor
Syntax
CURSor:VBArs:POSITION<x> <NR3>
CURSor:VBArs:POSITION<x>?
Related Commands
CURSor:VBArs:UNIts
Arguments
<NR3> species the cursor position.
Returns
A oating point value with an exponent.
Examples
CURSor:VBArs:POSITION2 9.00E-6 positions the cursor2 vertical bar cursor
at 9 ms.
CURSor:VBArs:POSITION1? this command might return
:CURSOR:VBARS:POSITION1 1.0000E-06 indicating that the
cursor1 vertical bar is positioned at 1 μs.
CURSor:VBArs:UNIts
Sets or returns the units for the vertical bar cursors.
Group
Cursor
Syntax
CURSor:VBArs:UNIts {SEConds|HERtz|D EGrees|PERcent}
CURSor:VBArs:UNIts?
Arguments
SEConds sets the units of the vertical bar cursors for the time domain (seconds).
HERtz sets the units of the vertical bar cursors for the frequency domain (Hertz).
DEGrees sets the units to degrees for use with an XY display.
PERcent sets the units to percent for use with ratio cursors.
MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual 2-117

Other manuals for Tektronix DPO2014

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO2014 and is the answer not in the manual?

Tektronix DPO2014 Specifications

General IconGeneral
BrandTektronix
ModelDPO2014
CategoryTest Equipment
LanguageEnglish

Related product manuals