EasyManuals Logo
Home>Tektronix>Test Equipment>DPO2024B

Tektronix DPO2024B Programmer's Manual

Tektronix DPO2024B
441 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #125 background imageLoading...
Page #125 background image
Commands Listed in Alphabetical Order
Syntax
CH<x>:SCAle <N
R3>
CH<x>:SCAle?
Related Commands
CH<x>:OFFSet, CH<x>:POSition, REF<x>:VERTical:SCAle,
MATH[1]:VERTical:SCAle
Arguments
<NR3> is the vertical channel scale in units-per-division. The value entered here
is truncated to three signicant digits.
Examples
CH4:SCALE 100E-03 sets the channel 4 scale to 100 mV per division.
CH2:SCALE? might return : CH2:SCALE 1.0000 indicating th
at the current scale
setting of c hannel 2 is 1 V per division.
CH<x>:TERmination
Sets the connected-disconnected status of a 50 resistor, which may be connected
between the specied channel's coupled input and o scilloscope ground. The
channel is specied by <x>. There is also a correspond
ing query that requests the
termination parameter and translates this enumeration into one of the two oat
values. This command is maintained for compatibility.
Group
Vertical
Syntax
CH<x>:TERmination
CH<x>:TERmination?
CH<x>:VOLts
Sets or returns the vertical sensitivity for channel <x>, where x is the channel
number.
Group
Vertical
Syntax
CH<x>:VOLts <NR3>
CH<x>:VOLts?
Arguments
<NR3> is the vertical sensitivity, in volts.
MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual 2-109

Other manuals for Tektronix DPO2024B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO2024B and is the answer not in the manual?

Tektronix DPO2024B Specifications

General IconGeneral
BrandTektronix
ModelDPO2024B
CategoryTest Equipment
LanguageEnglish

Related product manuals