Display a wavefo
rm
Setting the Display Persistence
Select Displa
y > Display Persistence, and
then select the type of persistence.
No persistence shows record points for
the current acquisition only. Each new
waveform rec
ord replaces the previously
acquired record for a channel.
Innite persistence continuously
accumulates record points until you
change one o
f the acquisition display
settings. Use for displaying points that
may occur outside the normal acquisition
envelope.
Variable p
ersistence accumulates record
points for a specied time interval.
Each record point decays independently
accordin
g to the time interv al.
Reset per
sistence clears the persistence.
1. To set the variable persistence time,
select D
isplay > Display Persistence >
Persistence Controls....
2. Click Variable, Persist Time, and then
use the multipurpose knobs to set the
persis
tence time.
92 MSO/DPO70000DX, MSO/DPO70000C, DPO7000C, and MSO/DPO 5000B Series U ser M anual

Table of Contents

Other manuals for Tektronix DPO71254C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO71254C and is the answer not in the manual?

Tektronix DPO71254C Specifications

General IconGeneral
BrandTektronix
ModelDPO71254C
CategoryTest Equipment
LanguageEnglish

Related product manuals