Commands Listed in Alphabetical Order
Group
Measurement
Syntax
MEASure[:SCALar]:POWer[:DC]? [CH1|CH2|CH3|ALL]
Related Commands
FETCh[:SCALar]:POWer[:DC]?, INSTrument:SELect
Arguments
CH1, CH 2, or CH3 is the channel on which to read the output power.
ALL is to read the output power on all channels.
Returns
<NR2> is the measured output power in watts.
Examples
MEAS:POW? ALL might return 9.97077, 0.00205158, 0 which would be the power
beingsuppliedonchannels1,2,and3inwatts.
MEASure[:SCALar][:VOLTage][:DC]? (Query Only)
This command initiates and executes a new voltage measurement, and returns the
measured output voltage of the power supply. If a channel is specified, the query
returns the measurement for the specified channel. If no channel is specified the
currently-selected c hannel is measured and returned.
Group
Measurement
Syntax
MEASure[:SCALar][:VOLTage][:DC]? [CH1|CH2|CH3|ALL]
Related Commands
FETCh[:SCALar]:VOLTage[:DC]?, INSTrument:SELect
Arguments
CH1, CH 2, or CH3 is the channel on which to read the output voltage.
ALL is to read the output voltage on all channels.
Returns
<NR2> is the measured output voltage in volts.
Examples
MEAS:VOLT? ALL might return 20.0002, 0.999465, 4.00024 which would be the
measured voltages on channels 1, 2, and 3 in volts.
Series 2200 Programmable Multichannel DC Power Supplies Programmer Manual 2-21