Commands Listed in Alphabetical Order
You can use commands to either set instrument features or query instrument
values. You can use some commands to do both, some only to set and some only
to query. Thi
s document marks set-only commands with the w ords “No Query
Form” included with the command name. It marks query-only commands with
a question mark appended to the header, and includes the words “Query Only”
in the command name.
This document spells out headers, mnemonics, and arguments with the minimal
spelling shown in uppercase. For example, to use the abbreviated form of the
MEASure:SCALar:VOLTage:DC? command, type MEAS:SCAL:VOLT:DC?.
*CLS (No Query Form)
The *CLS
command clears all event registers and queues.
Group
Status
Syntax
*CLS
Related Commands
*ESR?, *STB?
DISPlay[:WINDow][:STATe]
Sets or queries the state of the instrument display.
Group
Display
Syntax
DISPlay[:WINDow][:STATe] { 0|1|ON|OFF}
DISPlay[:WINDow][:STATe]?
Arguments
0|1|ON|OFF
DISPlay[:WINDow]:TEXT[:DATA]
Sets or queries the state of the instrument display to show a particular string.
Group
Display
Series 2200 Programmable Multichannel DC Power Supplies Programmer Manual 2-13