Commands Listed in Alphabetical Order
Arguments
0 disables the c
urrent channel.
1 enables the current channel.
Returns
Disabled
Enabled
[SOURce:]OUTPut:PARallel[:STATe]
This command sets the parallel state of CH1 and CH2.
Group
Source
Syntax
[SOURce:]OUTPut:PARallel[:STATe] 0|1|OFF|ON
[SOURce:]OUTPut:PARallel[:STATe]?
Arguments
0 or OFF sets the parallel state to off.
1 or ON sets the parallel state to on.
Examples
[SOURCE:]OUTPUT:PARALLEL[:STATE]
[SOURce:]OUTPut:PON[:STATe]
This command congures the power supply to power up with its output turned
off, or to return the output to the state it was in when it powered down.
Group
Source
Syntax
[SOURce:]OUTPut:PON[:STATe] {RST|RCL0}
[SOURc
e:]OUTPut:PON[:STATe]?
Arguments
RST sets the power supply to power-up with output off.
RCL0 sets the power supply to power-up with the output in the last state before
power was removed.
Returns
RST|RCL0
Series 2200 Programmable Multichannel DC Power Supplies Programmer Manual 2-29