EasyManuals Logo
Home>Tektronix>Test Equipment>MDO3014

Tektronix MDO3014 User Manual

Tektronix MDO3014
243 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #147 background imageLoading...
Page #147 background image
Analyze Wavefor
morTraceData
Selecting Automatic Measurem ents in the Time Domain
The following tables list each automatic measurement by category: time or amplitude. (See page 124, Taking A u tomatic
Measurements in the Time Domain.)
Time measurements
Measurement Description
Frequency
The rst cycle in a waveform or gated region. Frequency is the reciprocal of the period; it
is measured in hertz (Hz) where one Hz is one cycle per second.
Period
The time required to complete the rst cycle in a waveform or gated region. Period is
the recipro
cal of frequency and is measured in seconds.
Rise Time
The time required for the leading edge of the rst pulse in the waveform or gated region
to rise from the low reference value (default = 10%) to the high reference value (default =
90%) of the nal value.
Fall Time
The time required for the falling edge of the rst pulse in the waveform or gated region to
fall from the high reference value (default = 90%) to the low reference value (default =
10%) of the nal value.
Delay
The time between the mid reference (default 50%) amplitude point of two different
waveforms. See also Phase.
Phase
The amount of time that one waveform leads or lags another waveform, expressed in
degree
s where 360° makes up one waveform cycle. See also Delay.
Positive Pulse
Width
The distance (time) between the mid reference ( default 50%) amplitude points of a
positive pulse. The measurement is made on the rst pulse in the waveform or gated
region.
Negative Pulse
Width
The distance (time) between the mid reference ( default 50%) amplitude points of a
negative pulse. The measurement is made on the rst pulse in the waveform or gated
region.
Positive Duty
Cycle
The ratio of the positive pulse width to the signal period expressed as a percentage. The
duty cycle is measured on the rst cycle in the waveform or gated region.
Nega
tive Duty
Cycle
The ratio of the negative pulse width to t he signal period expressed as a percentage.
The duty cycle is measured on the rst cycle in the waveform or gated region.
Burst Width
The
duration of a burst (a series of transient events) and is measured over the entire
waveform or gated region.
MDO3000 Series Oscilloscopes User Manual 125

Table of Contents

Other manuals for Tektronix MDO3014

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO3014 and is the answer not in the manual?

Tektronix MDO3014 Specifications

General IconGeneral
BrandTektronix
ModelMDO3014
CategoryTest Equipment
LanguageEnglish

Related product manuals