Specifications
Table 4: Trigger specifications (cont.)
Characteristic Description
Standard serial bus interface triggering
I
2
C Address Triggering: 7 and 10 bit user specifi ed address, as well as General Call, START
byte, HS-mode, EEPROM, and CBUS
Data Trigger: 1 to 5 bytes of user specified data
Trigger O n: Start, Repeated Start, Stop, Missing Ack, Data, or Address and Data
Maximum Data Rate: 10 Mbps
SPI Data Trigger: 1 to 16 bytes of user specified data
Trigger On: SS Active, Start of Frame, MOSI, MISO, or MOSI and MISO
Maximum Data Rate: 10 Mbps
CAN Data Trigger: 1 to 8 bytes of user specified data, including qualifiers of equal to (=), not equal
to (<>), less than (<), greater than (>), less than or equal to (<=), greater than or equal to (>=)
Trigger On: Start of Frame, Type of Frame, Identifier, Data, Identifier and Data, E nd of Frame,
Missing Ack, or Bit Stuffing Error
Frame Type: Data, Remote, Error, Overload
Identifier: Standard (11 bit) and Extended (29 bit) identifiers
Maximum Data Rate: 1 M bps
RS-232 Trigger On: Tx Start Bit, Rx Start Bit, Tx End of Packet, Rx End of Packet, Tx Data, Rx Data,
Tx Parity Error, or Rx Parity Error
Maximum Data Rate: 10 Mbps
LIN
4
Identifier Trigger: 6 bits of user-specified data, equal to (=)
Data Trigger: 1 to 8 bytes of user-specified data, including qualifiers of equal to (=), not equal
to (<>), less than (<), greater than (>), less than or equal to (<=), greater than or equal to
(>=), inside range, or outside range
Error: Sync, Identifier Parity, Checksum
Trigger On: Sync, Identifier, Data, ID & Data, Wakeup Frame, S leep Frame, or Error
Maximum Data Rate: 100 kbps
MSO4000 and DPO4000 Series Specifications and Performance Verification 11