EasyManuals Logo
Home>Tektronix>Test Equipment>MSO54

Tektronix MSO54 Service Manual

Tektronix MSO54
50 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #20 background imageLoading...
Page #20 background image
Remove handle
Use this procedure to remove and replace the handle assemblies.
Before you begin
To prevent electrostatic damage to components whenever you work on the instrument, wear properly-grounded electrostatic prevention
wrist and foot straps, and work in a tested antistatic environment on an antistatic mat.
About this task
Remove the handle assemblies (table 6, index #13) using the following steps.
Procedure
1. Use a screwdriver with T-10 Torx tip to remove two screws (table 6, index #18) from each side of the handle.
2. Disengage the handle from the pins on the couplers (table 6, index #17) and remove the handle.
3. Remove the two couplers.
4. To reinstall, reverse the steps. Use a screwdriver with T-10 Torx tip to secure the screws. Tighten to 0.65 N·m.
Remove rear grill and case
Use this procedure to remove and replace the rear grill and rear case.
Before you begin
To prevent electrostatic damage to components whenever you work on the instrument, wear properly-grounded electrostatic prevention
wrist and foot straps, and work in a tested antistatic environment on an antistatic mat.
Remove handle on page 20.
Remove feet on page 19.
About this task
Remove the grill (table 6, index #14) and rear case (table 6, index #10) using the following steps:
Procedure
1. Use a screwdriver with T-8 Torx tip to remove the six screws securing the grill to the rear case (table 6, index #15) .
2. Put fingers in handle hub holes on each side of the grill and pull the grill off of the rear case.
3. Use a screwdriver with T-10 Torx tip to remove the six screws securing the rear case to the rear chassis (table 6, index #9) .
4. Slide the rear case off of the rear chassis.
5. To reinstall, reverse the steps. Tighten the six T-10 Torx rear case screws to 0.65 N·m. Tighten the six T-8 Torx rear grill screws to 0.45
N·m.
Remove rear chassis assembly
Use this procedure to remove and replace the rear chassis, to access rear chassis assemblies (power supply, main fan, Carrier interface
assembly, AFG Riser board).
Before you begin
To prevent electrostatic damage to components whenever you work on the instrument, wear properly-grounded electrostatic prevention
wrist and foot straps, and work in a tested antistatic environment on an antistatic mat.
Remove rear grill and case on page 20
CAUTION: Order T
ektronix part number 253-0582-XX (TAPE, ALUMINUM, BOTTOM PRE-CUT ) before removing the rear
chassis. The aluminum tape that is installed on the bottom front between the rear chassis and the front chassis cannot be reused.
The tape must be replaced after reinstalling the rear chassis assembly.
Maintenance
5 Series MSO Service Manual MSO54, MSO56, MSO58 20

Table of Contents

Other manuals for Tektronix MSO54

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO54 and is the answer not in the manual?

Tektronix MSO54 Specifications

General IconGeneral
BrandTektronix
ModelMSO54
CategoryTest Equipment
LanguageEnglish

Related product manuals