Removal and replace procedures
This section contains procedures for removal and installation of replaceable modules in the instrument. Refer to Replaceable
parts on page
35 for lists and exploded views of replaceable modules.
Any module inside of the chassis that does not have a remove and replace procedure requires that the entire instrument be returned to
Tektronix Service Center for service.
Warning: Before performing this or any other procedure in this manual, read the safety summaries found at the beginning of
this manual. Also, to prevent possible injury to service personnel or damage to the instrument components, read Preventing
ESD on page 15.
Before performing any procedure in this subsection, disconnect the power cord from the line voltage source. Failure to do so could
cause serious injury or death.
Before removing the rear chassis, order the 8 pieces of pre-cut aluminum tape (T
ektronix part number 253-0582-XX). The existing
tape has to be removed from the chassis to access the interior and cannot be re-applied after removal. The aluminum tape is only
needed for instruments with a serial number below B020000 or below C040000.
Note: Read the cleaning procedure before disassembling the instrument for cleaning.
Required equipment
Most assemblies in this instrument can be removed with a T
-10 or T8 Torx® screwdriver tip.
Table 3: Required equipment for removal and replacement
Item No. Name Description
1 Screwdriver handle Accepts Torx-driver bits
2 T-10 Torx tip Used for removing instrument screws. Torx-driver bit for T-10 size
screw heads
3 T-8 Torx tip Used for removing instrument screws. Torx-driver bit for T-8 size
screw heads
4 3/16 inch open-end wrench Used to remove nut posts
5 Needle-nosed pliers, small or medium size Used to pull off aluminum shielding tape from rear case and flat
cable.
6 Proper antistatic work environment To prevent electrostatic damage to components whenever you work
on the instrument, wear properly-grounded electrostatic prevention
wrist and foot straps, and work in a tested antistatic environment on
an antistatic mat.
Remove front-panel knobs
Use this procedure to remove and replace front-panel knob assemblies.
Before you begin
To prevent electrostatic damage to components whenever you work on the instrument, wear properly-grounded electrostatic prevention
wrist and foot straps, and work in a tested antistatic environment on an antistatic mat.
About this task
There are eight knob assemblies that can be removed from the front case.
Maintenance
5 Series MSO Service Manual MSO54, MSO56, MSO58 18

Table of Contents

Other manuals for Tektronix MSO56B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO56B and is the answer not in the manual?

Tektronix MSO56B Specifications

General IconGeneral
BrandTektronix
ModelMSO56B
CategoryTest Equipment
LanguageEnglish

Related product manuals