EasyManuals Logo
Home>Tektronix>Measuring Instruments>RSA5106B

Tektronix RSA5106B Technical Reference

Tektronix RSA5106B
153 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #138 background imageLoading...
Page #138 background image
Performance Verification
Spurious Resp
onse
Residual Response
1. Terminate the RSA5100B RF Input.
2. Reset the RSA5100B to factory defaults: select Setup > Preset (Main).
3. Select Tools > Alignments and select Align Now.
4. Set the RSA5100B:
Ref Level
Setup > Amplitude >
-40 dBm
Internal Attenuator
Setup > amplitude > Internal Settings > Internal
Attenuator
0 dB (Auto unchecked)
Setup > Amplitude > RF & IF Optimization >
Maximum Dynamic Range
RBW
Setup > Settings > BW > RBW
1kHz
Function
Setup > Settings > Traces > Function
Avg (VRMS)
Count
Setup > Settings > Traces > Function > Avg (of
logs)
10 (Count checked)
5. Set the Frequency (Freq key) to each center frequency in the Residual
Response Center Frequencies table. Set the span for each center frequency
as shown in the table. (See Table 80.) Wait for the averaging to complete,
then press the Peak key on the Markers control bar and record the marker
amplitude in the table.
6. Enter the highest of these signal levels into the test record.
Table 80: Residual response center frequencies
Center
frequency Span Range Marker amplitude
All models
600 MHz 800 MHz
200 MHz to 1 GHz
1.5GHz 1GHz 1GHzto2GHz
2.5GHz 1GHz 2GHzto3GHz
RSA5106B/5115B/5126B only
3.5GHz 1GHz 3GHzto4GHz
4.5GHz 1GHz 4GHzto5GHz
5.6 GHz 1.2 GHz 5 GHz to 6.2 GHz
RSA5115B/5126B only
124 RSA5100B Series Technical Reference

Table of Contents

Other manuals for Tektronix RSA5106B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix RSA5106B and is the answer not in the manual?

Tektronix RSA5106B Specifications

General IconGeneral
BrandTektronix
ModelRSA5106B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals