Specifications
Table 9 : Amplit
ude and RF atness (excluding mismatch error)
1
Characteristic Description
Reference level setting range, nominal –170 dBm to +40 dBm, 0.1 dB step, (Standard RF input)
Frequency response (18 °C to 28 °C)
10 MHz to 32 M
Hz, LF Band
±0.2 dB
10MHzto3GH
z
±0.35 dB
3GHzto6.2
GHz
(RSA5106B/5115B/5126B)
±0.5 dB
6.2 GHz to 15 GHz
(RSA5115B/5126B)
±1.0 dB
At 10 dB RF
attenuator setting
15 GHz to 26.5 GHz
(RSA512
6B)
±1.2 dB
10 MHz to 32 MHz, LF Band
(RSA510
3B/5106B)
±0.5 dB
10 MHz t
o3GHz
±0.5 dB
3GHzto
6.2 GHz
(RSA5106B/5115B/5126B)
±0.7 dB
6.2 GHz to 15 GHz
(RSA5115B/5126B)
±1.0 d
B
At 10 dB RF
attenuator setting,
With Preamp ON
15 GHz to 26.6 GHz
(RSA
5126B)
±1.2 dB
All RF attenuator
set
tings
1 Hz to 100 Hz, LF Band ±0.7 dB