TAS 200 Series Adjustment Procedures
Bench Test Instruments and Handheld Oscilloscopes Basic Service
23
Test Equipment
To ensure accurate adjustments, use the following or equivalent test equipment.
Alternative test equipment must meet or exceed the intended minimum
requirements specified in Table 19. If you substitute equipment, you may need to
modify the procedures.
NOTE. Before making any adjustment, warm up the test equipment according to
the manufacturer’s recommendations.
Table 19: Adjustment t est equipment
Description Minimum requirements Example product
Leveled Sine Wave Generator 1 kHz to 250 MHz; variable amplitude from
5mVto4V
p-p
into 50 Ω
Wavetek 9100 Universal Calibration
System with Oscilloscope Calibration
Time Mark Generator Variable marker frequency from 1 ms to
10 ns; accuracy within 2 ppm
Module (Opti on 250)
Fluke 5500
Multi-produ
tC
libr
tor with
DC Calibration Generator DC voltage level s f rom 100 mV to 10 V
Oscilloscope Calibration Option 5500A-SC
Digital Multimeter 35 V, 4 A, ±0.1% Accuracy Tektronix DMM914 Digital Multimeter
Termination (two required) 50 Ω, female BNC input, male BNC output Tektronix 011-0049-01
Cable, Coaxial 75 Ω, male-to-male BNC connect ors,
36 inch length
Tektronix 012-1338-00
Cable, Precisi on Coaxi al 50 Ω, male-to-male BNC connect ors,
36 inch length
Tektronix 012-0482-00
Coupler Female-BNC-to-dual-mal e-BNC, dual-input Tektronix 067-0525-02
Probe 10X probe Tektronix P6109B