EasyManua.ls Logo

Tektronix TAS 250 - Page 33

Tektronix TAS 250
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TAS 200 Series Adjustment Procedures
Bench Test Instruments and Handheld Oscilloscopes Basic Service
29
3. Rotate the front panel HORIZONTAL POSITION control to move the dot to
the display center. Rotate the INTENSITY control to give the dot a sharp
edge.
4. Adjust VR604 on the Power and High Voltage board and the oscilloscope
front panel FOC US control until the dot becomes circular. See Figure 7 for
the adjustment location.
5. Set the front panel HORIZONTAL X-Y push switch to the out position.
6. Repeat steps 1 through 5 for best performance.

Other manuals for Tektronix TAS 250

Related product manuals