Clear NAND Flash data using TekSecure
Procedure to clear NAND Flash data using TekSecure:
1. Before doing TekSecure, save your user data into a USB thumb drive. See
the following steps:
■
Press Save/Recall on the front panel.
■
Select Save Setup (bezel 1) , select Save to USB file (bezel 2), press
Save (bezel 5), to save setup to USB.
■
Select Save Waveform (bezel 1), select Save to USB file (bezel 2), press
Save (bezel 5), to save waveform to USB.
2. Enter into Utility > Config > press Teksecure Erase Memory.
3. There will be a pop up menu of warning showing: "This procedure will erase
all nonvolatile RAM, All reference waveforms and save setups will be
erased. Calibration constants will be retained. This will take approximately
3 minutes. Are you sure you want to erase?"
4. Press Yes. It will begin TekSecure, and after it finishes, it will reboot the
instrument.
Media and data export devices
The following table lists the data export devices in the instrument.
Clear and sanitize procedures
TBS2000B Series Declassification 5