EasyManuals Logo
Home>Tektronix>Test Equipment>TDS 684A

Tektronix TDS 684A Programming Manual

Tektronix TDS 684A
382 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #224 background imageLoading...
Page #224 background image
Command Descriptions
2–206
TDS Family Oscilloscope Programmer Manual
8 characters followed by a period (“.”) and any 3-character extension. Do not use
wild card characters.
RECALL:WAVEFORM TEK00000.WFM",REF1
recalls the waveform stored in the file named TEK00000.WFM to reference
location 1.
REM (No Query Form)
Specifies a comment. This line is ignored by the instrument.
Miscellaneous
REM <QString>
REM <Space>
<QString>
<QString> is a string that can have a maximum of 80 characters.
REM "This is a comment"
is ignored by the instrument.
*RST (No Query Form)
(Reset) returns the digitizing oscilloscope to a known set of instrument settings,
but does not purge any aliases or stored settings.
Status and Error
FACtory, *PSC, *RCL, RECAll:SETUp, *SAV, SAVe:SETUp
*RST
*RST
*RST returns the instrument settings to the factory defaults (see Appendix D:
Factory Initialization Settings).
Examples
Group
Syntax
Arguments
Examples
Group
Related Commands
Syntax

Table of Contents

Other manuals for Tektronix TDS 684A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS 684A and is the answer not in the manual?

Tektronix TDS 684A Specifications

General IconGeneral
BrandTektronix
ModelTDS 684A
CategoryTest Equipment
LanguageEnglish

Related product manuals