Application Examples
56
TDS1000/2000-Series Digital Oscilloscope User Manual
Capturing a Single-Shot S ignal
The reliability of a reed relay in a piece of equipment has been poor
and you need t o investigate the problem. You suspect that the relay
contacts arc when the relay ope ns. The fastest you can open and
close the relay is about once per mi nute so you need to capture the
voltage across the relay as a single-shot acquisition.
To set up for a singl e-shot acquisition, follow these steps:
1. Turn the vertical VOL TS/DIV and horizontal SEC/DIV knobs to
the appropriate ranges for the signal you expect to see.
2. Push the ACQUIRE button to see the Acquire Menu.
3. Push the Peak Detect option button.
4. Push the TRIG MENU button to see the Trigger Menu.
5. Push the Slope option button and select Rising.
6. Turn the LEVEL knob to adjust the trigger level to a voltage
midway between the open and closed voltages of the relay.
7. Push the SINGLE SEQ button to start the acquisition.
When the relay opens, the oscilloscope triggers and captures the
event.

Table of Contents

Other manuals for Tektronix TDS1000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS1000 Series and is the answer not in the manual?

Tektronix TDS1000 Series Specifications

General IconGeneral
BrandTektronix
ModelTDS1000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals