Maintenance
Probe Comp Output
When the oscill
oscope is on and operating properly, the Probe Comp output
should generate a square wave, approximately 5 V in amplitude, at a 1 kHz
frequency. Use the oscilloscope and use a 10X TPP0101 or TPP0201 probe to
probe this output.
As shown in the following table, there are two cases of defects. Either the 1 kHz
signal is on and the power supply and Main board are both active and functionally
capable of d
oing acquisition, or one of those two modules is not performing
correctly. While many different possible frequencies exist, all involve detected
failures on the Main board due to Main board failure or power supply failure.
Table 6-4: Probe Comp output
Signal at P robe Comp Possible problem
1 k Hz signal, no display
Refer to Troubleshooting the Display. (See page 6-30,
Troubleshooting the Display.)
Non-1 kHz signal
Refer to Troubleshooting the Main Board. (See page 6-37,
Troubleshooting the Main B oard .)
No signal
Refer to Troubleshooting the Power Supply. (See page 6-29,
Troubleshooting the Power Supply.)
6–28 TDS2000C Series Oscilloscope Service Manual

Table of Contents

Other manuals for Tektronix TDS2012C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS2012C and is the answer not in the manual?

Tektronix TDS2012C Specifications

General IconGeneral
BrandTektronix
ModelTDS2012C
CategoryOscilloscope
LanguageEnglish

Related product manuals