Specifications
Table 10: Data l
ogging system characteristics (not available on T DS1000C-EDU models)
Characteristic Description
Duration The time perio
d.
30 min, 1 hour, 1 .5 hour, 2 hour, 2.5 hour, 3 hour, 3.5 hour, 4 hour, 4.5 hour, 5 hour, 5.5 hour, 6 hour, 6.5
hour, 7 hour, 7.5 hour, 8 hour.
Source The signal source which you want to save the waveform.
Channel 1, C
hannel 2, Channel 3, Channel 4, Math
Select Fold
er
The file fold
er where you save the waveform data.
You can create the new folder or change the existing folder as the folder where you want to save the
waveform data.
Limit Testing System Characteristics (Not Available on TDS1000C-EDU Models)
NOTE. This software feature directs the oscilloscope to monitor an active input
signal against a template and to output pass or fail results by judging whether
the input signal is within the bounds of the template.
Table 11: Limit testing sy stem chara cteristics (not available on TDS1000C-EDU models)
Charac
teristic
Descri
ption
Sourc
e
The signal source which you want to do the limit testing.
Chann
el 1, Channel 2, Channel 3, Channel 4, Math
Compa
re Ref Channel
The re
ference channel # where the template is saved.
RefA, RefB, RefC, RefD. The limit testing system w ill compare the source signal with this template.
Run/Stop To enable or disable the limit testing function.
Run, Stop.
Template Setup You can use this menu item to set up a limit test waveform template. The template is the mask signal that
you
define as the boundary to compare with the input source signal. You can create the template from
internal or external waveforms with specific horizontal and vertical tolerances.
Source The location of the signal source that is used to create the limit test template.
Channel 1, Channel 2, Channel 3, Channel 4, Math
Vertical Limit The vertical lim it in vertical divisions.
0~1000 mdiv
Horizontal Limit The horizontal limit in horizontal divisions.
0~500 mdiv.
Destination Ref Channel The location of the reference memory location that is used to store the limit test template.
RefA, RefB, RefC, RefD.
Display Template Displays or does not display a stored test template.
On, O ff.
16 TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual