Acquisition
How the Acquisition Modes Work
Sample mode retains the rst sampled point
from each acquisition interval. Sample is the
default mode.
Peak Detect mode uses the highest and
lowest of all the samples contained in two
consecutive acquisition intervals. This mode
only w orks with real-time, noninterpolated
sampling and is useful for catching high
frequency glitches.
Hi Res mode calculates the average
of all the samples for each acquisition
interval.
Hi-Res provides a higher-resolution,
lower-bandwidth waveform.
Envelope mode nds the highest and
lowest record points over many acquisitions.
Envelope u
ses Peak Detect for each
individual acquisition.
Average mode calculates the average value
for each record point over many acquisitions.
Average uses Sample mode for each
individual acquisition. Use average mode to
reduce random noise.
Waveform Database mode is a
three-dimensional accumulation of source
waveform data over several acquisitions. In
addition to amplitude and timing information,
the database includes a count of the number
of times a specic waveform point (time and
amplitude) has been acquired.
TDS6000B & TDS6000C Series Quick Start User Manual 21

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS6000B Series and is the answer not in the manual?

Tektronix TDS6000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS6000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals