State Acquisition
You can use the logic analyzer for state data analysis; you can view the data in tabular form in the Listing window.
Define the Setup Window
State acquisition uses an external clock from the target system to tell the logic analyzer when to sample data.
1. Configure the signals in the Setup
window. (See page 8, Configure the Setup
Window.)
2. Select Synchronous.
3. Select the Single Clock source. For this
example, select CK0().
4. Select the rising edge of the clock.
5. Select Samples.
Configure the Trigger
1. Define the setups for the Waveform
window. (See page 15, Define the
Waveform Window.)
You will use the Waveform window
setups to help de fine the trigger.
2. Click Trigger in the tabbed window.
3. Click and drag the Channel Edge icon
from the Tabbed Trigger window to the
Output waveform.