EasyManuals Logo
Home>Thermo>Chiller>NESLAB RTE Series

Thermo NESLAB RTE Series User Manual

Thermo NESLAB RTE Series
52 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #48 background imageLoading...
Page #48 background image
The master sets parameters in the bath by sending one of the Set Functions as shown in Table 1. The master
does not send a qualifier byte in the data field. The master should be pre-programmed to send the correct
precision and units (it could also read the parameter of interest first to decode the correct precision and units
needed).
As an RS-485 example, if the master wants to set the setpoint to 30°C, it would send :
The bath responds:
command byte
2 bytes to follow
The checksum is the bitwise inversion of
22 (00+03+F0+02+01+2C)
The temperature of 30.0°C is 300 decimal = 012C hex.
CC 00 03 F0 02 01 2C DD
CC 00 03 F0 03 11 01 2C CB
command byte
3 bytes to follow
The checksum is the bitwise inversion
of 34(00+03+F0+03+11+01+2C)
The qualifier byte of 11 indicates a precision of 1
decimal point and units of °C. The temperature of
30.0°C is 300 decimal = 012C hex.
Appendix B
B - 3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo NESLAB RTE Series and is the answer not in the manual?

Thermo NESLAB RTE Series Specifications

General IconGeneral
BrandThermo
ModelNESLAB RTE Series
CategoryChiller
LanguageEnglish

Related product manuals