Table 2 Read Status
BIT d1 d2 d3
b.7 = 1 RTD1 Open Fault RTD2 Open Fault High Fixed Temp Fault
b.6 = 1 RTD1 Shorted Fault RTD2 Shorted Fault Low Fixed Temp Fault
b.5 = 1 RTD1 Open RTD2 Open Warn High Temp Fault
b.4 = 1 RTD1 Shorted RTD2 Shorted Warn Low Temp Fault
b.3 = 1 RTD3 Open Fault RTD2 Open Low Level Fault
b.2 = 1 RTD3 Shorted Fault RTD2 Shorted High Temp Warn
b.1 = 1 RTD3 Open Refrig High Temp Low Temp Warn
b.0 = 1 RTD3 Shorted HTC Fault Low Level Warn
BIT d4 d5
b.7 = 1 Buzzer On RTD2 Controlling
b.6 = 1 Alarm Muted Heat LED Flashing**
b.5 = 1 Unit Faulted Heat LED On**
b.4 = 1 Unit Stopping Cool LED Flashing**
b.3 = 1 Unit On Col LED On**
b.2 = 1 Pump On 0
b.1 = 1 Compressor On 0
b.0 = 1 Heater On 0
**LED bits in REQ STATUS1 byte d5
LED State LED Flashing LED On
OFF 0 0
ON 0 1
FLASHING 1 1
Invalid 1 0
b.7 = most significan bit
b.0 = least signifcant bit
Table 3 Qualifier Byte
10 hex 0.1 precision, no units of measure
20 hex 0.01 precision, no units of measure
11 hex 0.1 precision, °C units
21 hex 0.01 precision, °C units
Example: The temperature of 45.6 °C would be represented by the qualifier 11 hex, followed by the 2 bytes 01
C8 hex (456 decimal).
Appendix B
B - 5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo NESLAB RTE Series and is the answer not in the manual?

Thermo NESLAB RTE Series Specifications

General IconGeneral
BrandThermo
ModelNESLAB RTE Series
CategoryChiller
LanguageEnglish

Related product manuals