Page #68 background image
62
?
This is a ping command used to tell if the flowmeter is
communicating. The flowmeter will return an acknowledge sequence
of “OK” <CR> <LF> as a response.
Example 1) ? Returns “OK” <CR> <LF>
Error Codes
1 Unrecognizable commandThe flowmeter uses the length
of the command and the first few letters (how many letters
depends on the command) to recognize a valid command.
2 Number out of rangeThe number entered as the operand
to a command was out of the specified range or
unrecognizable.
3 Invalid modeOne or more requested options to a
command were invalid.
4 Command not possibleThe supplied operands describe a
command that is beyond the functional capability of the
flowmeter.
8 Internal errorAn internal failure was detected.
Question and Answer IconNeed help?

Do you have a question about the TSI Instruments 4000 Series and is the answer not in the manual?

TSI Instruments 4000 Series Specifications

General IconGeneral
Measurement ParametersVelocity, Temperature
Data Storage10, 000 data points
Accuracy±0.015 m/s or ±1% of reading
Particle Size RangeNot applicable
DisplayLCD
Operating Humidity0 to 95% RH (non-condensing)

Summary

Chapter 1 - Models and Accessories

Model Designations

Details specific model numbers, gas calibration, flow range, and adapter types.

Chapter 2 - Measurements Available

Flow Rate Measurement

Explains how flow rate data is obtained via analog or serial interfaces.

Chapter 4 - Design Considerations, Mechanical

Important Safety Notes

Covers hydrogen safety and intrinsic safety aspects of the flowmeters.

Chapter 5 - Design Considerations, Electrical

Power Requirements

Specifies voltage, current, and connection type for the flowmeter's power supply.

Output Signal, Analog

Describes the default analog output voltage range for mass flow measurement.

Chapter 7 - Specifications

Flow Measurement

Details measurement ranges and accuracy for different models and gases.

Accuracy (continued)

Provides accuracy specifications for Nitrogen and Nitrous Oxide gases.

Appendix C - Digital Serial Interface

Command Set Summary

Overview of data format, buffering, and command structure for serial communication.

Commands for Flow Rate, Volume, and Temperature Measurement

Details commands for acquiring flow rate, volume, and temperature measurement data.