11
MODEL RVA801/LCA301
SPECIFICATIONS
(subject to change without notice)
Measurement Range
Velocity 50 to 6000 fpm 0.25 to 30 m/s
Temperature 32 to 140 °F 0 to 60 °C
Volumetric
Flow Rate
1-9999000 cfm 1 - 999999 l/s
1 - 999999 m
3
/h
0.01 – 3000 m
3
/s
Area Input 0.043 - 900 ft
2
0.00399 - 90 m
2
Actual range is a function of velocity and area input.
Display Resolution
Velocity 1 fpm 0.01 m/s
Temperature 1 °F 0.1 °C
Accuracy
Velocity ±1% reading ±4 fpm ±1% reading ±0.02
m/s
Temperature ±2 °F ±1 °C
Storage
Temperature
14 to 140 °F -10 to 60 °C
Operating
Temperature
32 to 140 °F 0 to 60 °C
Power Source
9-volt battery
Battery Life
Approximately 40 hours continuous use
Dimensions
(LWD)
11 in. × 4.5 in. ×
2.6 in.
280 mm × 112 mm ×
65 mm
Weight
10 oz. 275 g
(excluding battery)