Service Manual UPO1000X Series
Instruments.uni-trend.com
12
/
31
Removal and Replacement
Remove Front Panel
The following process is describes the removal and replacement of knob module on the front panel. Seven knob
modules can be remove from the front panel.
Precondition
Whenever you work on the insturment, you should wear grounded anti-static wristbands and foot bands, and use
anti-static pads in a tested anti-static environment.
Steps
1. Pull out the knob module directly from the front shroud by hands. Hands force is enough to remove the knobs
module. Do not use pliers.
2. Align the knobs to axle indentation for reinstallation. Rotate the knobs to ensure the knobs is properly reinstalled
and can be rotate smoothly.
Remove Rear Panel
The following process is describes the removal and replacement of rear panel.
Precondition
To prevent static damage to the components while working on the instrument, wear properly grounded anti-static
wristbands and foot bands, and use anti-static pads in a tested anti-static environment.
Steps
1. Dismantle gum.