EasyManuals Logo
Home>UNI-T>Multimeter>UT71E

UNI-T UT71E User Manual

UNI-T UT71E
73 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #56 background imageLoading...
Page #56 background image
53
Model UT71A/B/C/D/E: OPERATING MANUAL
Warning
To avoid electrical shock or arc blast, or personal
injury or damage to the Meter, use specified fuses
ONLY in accordance with the following procedure.
Turn the rotary switch to OFF and remove all
connections from the terminals.
Remove the 5 screws from the case bottom.
Remove the fuse by gently prying one end loose,
then take out the fuse from its bracket.
Install ONLY replacement fuses with the identical
type and specification as follows and make sure the
fuse is fixed firmly in the bracket.
Fuse 1: 0.5A, 1000V, fast type fuse, 6.3x32mm
Fuse 2: 10A, 1000V, fast type fuse, 10.3x38mm
Rejoin the case bottom and case top, and install the
5 screws.
l
l
l
l
l
B. Replacing the Fuses
Figure 6-1. Fuse Replacement
Follow Figure 6-1 and proceed as follows to replace
the Meter's fuse:
Replacement of the fuses is seldom required. Burning
of a fuse always results from improper operation.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the UNI-T UT71E and is the answer not in the manual?

UNI-T UT71E Specifications

General IconGeneral
Data holdYes
Product typeDigital multimeter
Frequency range0.04 - 40000 kHz
AC current range0.4 - 10 A
AC voltage range4 - 1000 V
DC voltage range0.4 - 1000 V
Capacitance range40 - 40000 nF
Statistic functionMaximum, Minimum
Basic accuracy (frequency)0.01%+8
Basic accuracy (DC current)0.08%+8
Basic accuracy (DC voltage)0.025%+5
Basic accuracy (resistance)0.3%+8
Basic accuracy (capacitance)1%+20
Temperature measurement range-40 - 1000 °C
Dimensions (WxDxH)93 x 40 x 200 mm
Battery voltage9 V
Weight and Dimensions IconWeight and Dimensions
Weight384 g