EasyManuals Logo
Home>ABB>Transmitter>4600 Series

ABB 4600 Series User Manual

ABB 4600 Series
28 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #14 background imageLoading...
Page #14 background image
12
6.1 Instrument Start-up
Ensure all electrical connections have been made correctly and apply power to the instrument.
6.2 Operating Page
The operating page is a general use page in which continuously updated measured values and preset parameters can be viewed
but not altered. To adjust or set a parameter refer to the programming pages in Section 7.
6 OPERATION
% Oxygen
The upper display indicates the measured oxygen value. If over or under
temperature is monitored, the upper display shows '------' and an error
message is displayed on the lower dot matrix display – see Section 6.3.
The lower display indicates the measured cell temperature in °C.
Cell Temperature (°C)
The upper display indicates the measured cell temperature in °C.
Cell Millivolts
The upper display indicates the measured cell millivolts.
Alarm 1 Set Point
The upper display indicates the alarm 1 set point, displayed as % oxygen. The
set point value and the relay/l.e.d. action is programmed in the
Set Up Outputs
Page – see Section 7.7.
Alarm 2 Set Point
Note. This frame is not displayed if the Alarm 2 Action parameter has
been set to General Alarm – see Section 7.7.
The upper display indicates the Alarm 2 set point, displayed as % oxygen.
Press to return to the top of the Operating Page.
Press to advance to the beginning of the Oxygen Calibration Sequence.
Alarm 2 Setpoint
Alarm 1 Setpoint
20.00
5.00
0.0
Cell milliVolts
20.9
% Oxygen 700°C
700
Cell Temperature
OXYGEN CAL.
Note. If Alarm 2 has been programmed as a general system/instrument alarm, the associated front panel l.e.d. is
illuminated when the alarm is active and Relay 2 has de-energized.
6.3 Operating Page Error Messages
If an error is detected, the appropriate error message (see table below) will replace the % Oxygen frame in the Operating Page.
egasseMrorrE esuaCelbissoP
RORREYROMEMVN
ehtyfitceroT.purewopgnirudyltcerrocdaerneebtonevahyromemelitalov-nonehtfostnetnocehT
.ynapmoCehttcatnocstsisreptluafehtfI.niaganohctiwsdnasdnoces01tiaw,ffohctiws,tluaf
PMETREDNULLEC
006<(sgnidaerelbatiusniatbooterutarepmettneiciffusadehcaertonsaheborpehT ° .)C
LIAFNOITARBILAC
.deliafnoitarbilactniop-owtro-elgnistsalehT
TIUCRICNEPOC/T
sierutarepmetelpuocomrehtehtrotiucricnepoerasnoitcennocelpuocomrehtehT
0021nahtretaerg ° 2912/C ° 0051ro)elpuocomrehtKepyt(F ° 2372/C ° .)selpuocomrehtB&S,Repyt(F

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ABB 4600 Series and is the answer not in the manual?

ABB 4600 Series Specifications

General IconGeneral
BrandABB
Model4600 Series
CategoryTransmitter
LanguageEnglish

Related product manuals