EasyManuals Logo
Home>ABB>Transmitter>4600 Series

ABB 4600 Series User Manual

ABB 4600 Series
28 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #15 background imageLoading...
Page #15 background image
13
7 PROGRAMMING
Operating
Parameters
Secure
Parameters
20.9
% Oxygen
700
Cell Temperature
0.0
Cell milliVolts
20.00
Alarm 1 Setpoint
5.00
Alarm 2 Setpoint
Section 6.2
-----
OXYGEN CAL. SEQ.
0.0
Cell Constant mV
100.0
Span % of Theory
00000
Cal. User Code
Preset Cal.
One Point Cal.
-----
Two Point Cal.
0.0
Adjust Cell Zero
-----
Span Theory NO
YES
Section 7.1
Section 7.3
20.95
00000
Security Code
Connect to Air
20.95
Calibrating Air
-----
Calibration Pass
Failed Constant
Failed Unstable
20.95
Connect to Air
20.95
Calibrating Air
1.00
Enter Span Gas
1.00
Connect Span Gas
1.00
Calibrating Span
-----
Calibration Pass
Failed Span
Failed Constant
Failed Unstable
Section 7.4
Section 7.1
Section 7.5 Section 7.7
YES
Section A.3
Section 7.6
Section 7.2
xxxxx
English
-----
Deutsch
Français
Español
-----
SET UP OUTPUTS
-----
A1 Action EA
-----
A1 Setpoint
0.0
A2 Action EA
20.00
A2 Setpoint
-----
RTX Type 4-20
-----
Log Output NO
25.00
Retrans Span
0.0
Retrans Zero
-----
Hold Outputs NO
EB
EB
General Alarm
0-20
0-10
YES
YES
-----
ELECTRICAL CAL.
-----
Calibrate YES
xxxxx
mV Zero 1 -20mV
mV Span 1 180mV
xxxxx
mV Zero 2 10mV
xxxxx
mV Span 2 40mV
100.0
Test Retrans (%)
00000
Alter Sec. Code
00000
Alter Cal. Code
25.0
Adjust CJ Value
-----
Adjust RTX Zero
-----
Adjust RTX Span
K
-----
SET UP T/C
T/C TYPE NONE
R
S
-----
700
Preset Temp
NO
B
Fig. 7.1 Overall Programming Chart

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ABB 4600 Series and is the answer not in the manual?

ABB 4600 Series Specifications

General IconGeneral
BrandABB
Model4600 Series
CategoryTransmitter
LanguageEnglish

Related product manuals