SECTION 5
14
i-STAT
Alinity — Quick Reference Guide Art: 731848-01 Rev K Rev. Date: 02-Apr-2021
CLEANING
It is recommended that the i-STAT Alinity, base station and electronic
simulator be cleaned periodically or whenever visibly soiled.
Standard precautions should be taken whenever working with blood
or blood products.
SENSITIVE AREAS
When cleaning the i-STAT Alinity with CaviWipes, power off the
instrument and place it on a level surface. Do not clean or disinfect
the instrument while it is in the Base Station. The Base Station nor the
printer need to be unplugged when being cleaned.
Remove a new disposable wipe from the container and squeeze
to remove excess solution.
Gently wipe all outside surfaces (noting the “Sensitive Areas”) until
all visible soil is removed.
Inspect all surfaces. If necessary, repeat until all visible soil
is removed.
Wipe with dry gauze until dry.
1
2
3
4
5
i-STAT Alinity Instrument, Base Station, Printer and Electronic Simulator
Avoid forcing liquid into these areas:
i-STAT Alinity Instrument
Cartridge Port
10-Pin Connector under the camera
Gold contacts (2) on the outside of
the battery
A
B
C
D
E
F
Base Station
10-Pin Connector
Gold contact pins (2)
USB Port
D
E
F
Electronic
Simulator
Area between
protective cap
retaining ring
and white
sensor area
G
C
B
A