14 Agilent 8494/95/96G/H Attenuators Operating and Service Manual
1 Introduction
Instrument Options
Each instrument is specified with an option number which denotes
the configuration of the input and output connectors.
Do not exceed the RF power rating of 1 W average or 100 W peak with a
maximum pulse width of 10 μs. Do not connect an attenuator RF input or
output connector to greater than ±7 Vdc. If the attenuator must be connected
to a device with a potential greater than ±7 Vdc, use a blocking capacitor.
Table 4 Instrument Options
Option Description
001 Type-N (f) connectors
002 SMA female connectors
003
[1]
[1] Option 003 is not available on the 8495G.
APC-7 connectors
024
[2]
[2] Option 024 and 060 are default options unless specified otherwise.
24 Vdc supply voltage
011 5 V operation instead of standard 24 V
016 14-pin dip plug, 16-inch DC ribbon cable
060
[2]
Viking to bare wire cable