dismiss the Message Box. The fix was to re- enable the touchscreen before
displaying the Message box and the disabling it after it has been dismissed.
• If a vertical calibration fails and the module is removed before dismissing the
Message Box an illegal operation error occurs. The Fix is to check for the presence
of the module before reading from it.
• Timebase resolution changed from 3 digits to 4, giving the same behavior as the
83480A.
• Remotely invoking an extinction ratio cal, where a vertical cal is required, brings
up a dialog box that cannot be cleared. The dialog box is replaced with a status
message in the status bar.
• Remotely sending the TEDGE command caused various problems. The fix is to
only allow TEDGE? query and to reject TEDGE command.
• Remotely selecting vertical markers from a time waveform to a versus waveform
caused a "data out of range" error message; this has been fixed.
• Invalid dates remotely sent to the instrument cause several problems, which are
now resolved.
A.01.13 Release Notes, May 2, 2000 (Differences from Release
A.01.12)
NOTE This firmware release is not supported.
New Feature List
•None
Defects Fixed
• User Vertical Calibration routine that causes vertical calibration failures