N9310A RF Signal generator
Quick Start Guide 3
Safety Notice
Please read the following warnings and cautions carefully
before you power on the signal generator to ensure your
personal and instrumental safety.
WARNING
Always use a well-grounded, three-pin AC plug and power cord to
connect to a power source. Personal injury may occur if there is any
interruption of the AC power cord of the signal generator. Intentional
interruption is prohibited.
WARNING
Personal injury may result if the signal generator covers are removed.
There are no operator serviceable parts inside. To avoid electrical
shock, refer servicing to qualified personnel.
WARNING
Electrical shock may result if the signal generator is connected from
the
power supply while cleaning. Do not attempt to clean
internally.
CAUTION
To install the signal generators in other racks, note that they may
promote shock hazards, overheating, dusting contamination, and
inferior system performance. Consult your Agilent customer engineer
about installation, warranty, and support details.
CAUTION
Damage to the signal generator may result when the total power
dissipated in the cabinet is greater than 800 watts. When this
condition exists, forced convection must be applied.
CAUTION
The RF OUT connector is for signal output only. Avoid manually adding
any external signal into the signal generator via this connector. This
connector endures maximum +36 dBm RF power or 30 V DC input
(1 minute lasting). Or it may result in instrument damages.