1 Calibration Procedures
Recommended Test Equipment
4 U1231A/U1232A/U1233A Service Guide
Recommended Test Equipment
The test equipment recommended for the performance
verification and adjustment procedures is listed below in
Table 1- 1. If the exact instrument is not available, substitute
with another calibration standard of equivalent accuracy.
Table 1 - 1 Recommended test equipment
Application
Recommended
equipment
Recommended accuracy requirements
DC voltage Fluke 5520A
<20% of the instrument accuracy
specification
DC current Fluke 5520A
<20% of the instrument accuracy
specification
Resistance Fluke 5520A
<20% of the instrument accuracy
specification
AC voltage Fluke 5520A
<20% of the instrument accuracy
specification
AC current Fluke 5520A
<20% of the instrument accuracy
specification
Frequency Fluke 5520A
<20% of the instrument accuracy
specification
Capacitance Fluke 5520A
<20% of the instrument accuracy
specification
Diode Fluke 5520A
<20% of the instrument accuracy
specification
Temperature Fluke 5520A
<20% of the instrument accuracy
specification
Short
Shorting plug — a dual banana plug with a copper wire
shorting the two terminals

Table of Contents

Other manuals for Agilent Technologies U1231A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U1231A and is the answer not in the manual?

Agilent Technologies U1231A Specifications

General IconGeneral
BrandAgilent Technologies
ModelU1231A
CategoryMultimeter
LanguageEnglish

Related product manuals