G2S BPAP Series User Manual V2.1
5 / 43
Heated Humidifier
Humidifier Settings: Off, 1 to 5 (95˚F to 154.4˚F / 35°C to 68°C)
Humidifier Output: No less than 10 mg H
2
O/L
Environmental Conditions: Maximum airflow, 35°C, 15% relative humidity
Maximum Delivered Gas Temperature: 43°C
Mode of Operation
Continuous
Work Mode
CPAP, S, AutoS, AutoCPAP, S/T, T
SD Card
SD card can record patient data and fault information.
AC Power Consumption
100 - 240 V , 50 / 60 Hz, Max 2 A
Main Device offer to USB Communications Port
5 V 2.0 A
Type of Protection against Electric Shock
Class II Equipment
Degree of Protection against Electric Shock
Type BF Applied Part
Degree of Protection against Ingress of Water
IP22
Pressure Range
IPAP: 4.0 20.0 hPa (only applies to G2S B20A, G2S B20S, G2S B20T);
4.0 25.0 hPa (only applies to G2S B25A, G2S B25S, G2S B25T, G2S B25VT, G2S LAB);
4.0 30.0 hPa (only applies to G2S B30T, G2S B30VT, G2S B30AT);
in 0.5 hPa increments.
EPAP: 4.0 20.0 hPa (only applies to G2S B20A, G2S B20S, G2S B20T);
4.0 25.0 hPa (only applies to G2S B25A, G2S B25S, G2S B25T, G2S B25VT, G2S
B30T, G2S B30VT, G2S B30AT, G2S LAB);
in 0.5 hPa increments.
CPAP and AutoCPAP mode: 4.0 20.0 hPa
Under single fault conditions, 30 hPa for CPAP mode, 40 hPa for the rest modes.
Pressure Display Accuracy
± (0.8 hPa4%)
Static Pressure Stability
± 0.5 hPa
Ramp
The ramp time ranges from 0 to 60 minutes.