G2S A20 / G2S C20 User Manual V1.0
38 / 39
Recommended separation distances between RF wireless communications
equipment
The device is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the device can help prevent
electromagnetic interference by maintaining a minimum distance between RF wireless
communications equipment and the device as recommended below, according to the maximum
output power of the communications equipment.
Frequency
MHz
Maximum
Power
W
Distance
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment - Guidance
385
1.8
0.3
27
27
RF wireless communications
equipment should be used
no closer to any part of the
device, including cables,
than the recommended
separation distance
calculated from the equation
applicable to the frequency
of the transmitter.
Recommended
separation distance
Where
P
is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and
d
is the
recommended separation
distance in meters (m).
Field strengths from fixed
RF transmitter, as
determined by an
electromagnetic site
survey, should be less than
the compliance level in
each frequency range.
Interference may occur in
the vicinity of equipment
marked with the following
symbol:
450
2
0.3
28
28
710
0.2
0.3
9
9
745
780
810
2
0.3
28
28
870
930
1720
2
0.3
28
28
1845
1970
2450
2
0.3
28
28
5240
0.2
0.3
9
9
5500
5785
Note 1: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
WARNINGS!
This device should not be used in the vicinity or on the top of other electronic equipment
such as cell phone, transceiver or radio control products. If you have to do so, the device
should be observed to verify normal operation.
The use of accessories and power cord other than those specified, with the exception of
cables sold by the manufacturer of the equipment or system as replacement parts for
internal components, may result in increased emissions or decreased immunity of the
equipment or system.

Table of Contents

Other manuals for BMC G2S C20

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the BMC G2S C20 and is the answer not in the manual?

BMC G2S C20 Specifications

General IconGeneral
BrandBMC
ModelG2S C20
CategoryMedical Equipment
LanguageEnglish

Related product manuals