G2S A20 / G2S C20 User Manual V1.0
5 / 39
5. Specifications
Device Size
Dimensions: 274 mm × 184 mm × 115 mm
Weight: 1.9 kg
Water capacity:
To maximum fill line 360 mL
Product Use, Transport and Storage
Operation Transport and Storage
Temperature: 5°C to 35°C (41˚F to 95˚F) -25°C to 70°C (-13˚F to 158˚F)
Humidity: 15% to 93% Non-condensing 15% to 93% Non-condensing
Atmospheric Pressure: 760 ~ 1060 hPa 760 ~ 1060 hPa
Heated Humidifier
Humidifier Settings: off, 1 to 5 (95˚F to 154.4˚F / 35°C to 68°C)
Humidifier Output: No less than 10 mg H
2
O/L
Environmental Conditions: Maximum airflow, 35°C, 15% relative humidity
Mode of Operation
Continuous
Work Mode
CPAP, Auto
SD Card
The SD card can record patient data and fault information
AC Power Consumption
100 - 240 V ~, 50 / 60 Hz, Max 2 A
Device offer to USB Communications Port
5 V 2.0 A
Type of Protection Against Electric Shock
Class II Equipment
Degree of Protection Against Electric Shock
Type BF Applied Part
Degree of Protection Against Ingress of Water
IP22
Pressure Range
4 to 20 hPa (in 0.5 hPa increments), ≤30 hPa under single fault conditions.
Static Pressure Stability
± 0.5 hPa
Ramp
The ramp time ranges from 0 to 60 minutes.