ACS 753, ACS 763, ACS 863 Service Manual
SP00D00609 2018-03-26Robert Bosch GmbH
55
Diagnostics and Testing
PROCEDURE ERROR/WARNING DESCRIPTION OF ERROR/
WARNING
Oil Inject Oil inject error. Inject: XXXYY 1. Check if enough oil is present in oil
bottle.
2. Check for oil inject tube kinks or
cuts.
3. Check for oil inject solenoid failure.
4. Check for Oil inject stuck check
valve.
Recover No pressure on inlets.
Recover Anyway?
Select OK (Yes) or ESC icon (No).
Service Vacuum Vacuum halted! The inlet
pressure is too high for
vacuum.
Select RETRY or ESC icon.
Set Tank Tare Invalid entry. Must be
between XXYY and XXYY.
Tank tare range is either too low or
too high. Select Retry to re-enter
tank tare value or ESC icon.
Slow Charge Charge stalled! Check
connections and valves.
XX.XXXYY of XX.XXXYY
charged.
Check panel valve position and
couplers. Select RETRY or ESC
icon.
Startup System Fault. RTC Fault. System missing date/time
information. Select ESC icon to run
RTC routine.
System Flush Insufcient refrigerant. 6.00kg
required for system ush.
Select ESC icon, then perform tank
ll.
System Flush Pressure too high for ush.
System must be recovered.
Select ESC icon, then perform
recovery.
Tank Fill Source tank is empty. Filled:
XX.XXYY.
Source tank emptied during the tank
ll process. Connect new source
tank, then select RETRY or ESC
icon.
Vacuum The inlet pressure is too
high for vacuum. Recover or
Cancel.
Run a recovery by selecting Recover
or ESC icon to quit.
Vacuum Vacuum Halted! The inlet
pressure is too high for
vacuum.
Select RETRY or ESC icon.
Vacuum Test Test failed! Select RETRY or ESC icon.
Recovery
Refrigerant purity acceptable
(if the refrigerant identier is
present)
Refrigerant analysis test passed:
refrigerant purity acceptable