12
Product specification
Type: No touch 3 in 1 thermometer (BNT100)
Measuring range: Body: 34.0 °C – 43.0 °C (93.2 °F – 109.4 °F)
Food/Bath: 0 °C – 100.0 °C (32.0 °F – 212.0 °F)
Measuring location: Forehead
Resolution: 0.1 °C (0.1 °F)
Laboratory accuracy
- Body temperature mode: ±0.2 °C for the range 35.0 °C – 42.0 °C
(±0.4 °F for 95.0 °F – 107.6 °F)
±0.3 °C (±0.5 °F) for the range outside
35.0 °C – 42.0 °C (95.0 °F – 107.6 °F)
- Food/bath temperature mode: ±2.0 °C for the range 0.0 °C – 100.0 °C
(non-medical use) (±3.6 °F for 32.0 °F – 212.0 °F)
Clinical accuracy: Clinical performance and protocols are available upon request
This thermometer displays a calculated oral equivalent estimate
Display: Liquid Crystal Display, 4 digits plus special icons
Acoustic: Audio:
Normal temp range = Green temp range: 1 long beep for 0.4
second duration
Fever = Yellow temp range: 1 long beep for 0.4 second duration
High fever = Red: 10 short beeps (0.02 seconds on, 0.13 seconds o)
Operating temperature: 15 °C – 40 °C (59.0 °F to 104 °F)
Automatic Switch-o: Within approx. 20 seconds
Weight: 93.2g (w/ batteries) and 71.5 (w/o batteries)
Size: 55mm x 38mm x 149mm
Long term storage ranges
Storage/transport temperature: -25 °C to 55 °C (-13 °F to 131 °F )
Humidity: 15–95% non-condensing
Battery: (2) AAA Batteries - at least 500 measurements
Pressure: 700-1060 hPA (0.7-1.06 atm)
Memory size: 10
The expected service life of the thermometer is 5 years.
If necessary, the circuit diagram and component list can be provided for Authorized
Technician Maintenance.
The BNT100US includes the following items: No touch thermometer,
2 AAA batteries, owner’s manual and quick reference guide.
This infrared thermometer meets requirements established in ASTM Standard
E 1965-98 (for the thermometer system). Full responsibility for the conformance of
the product to the standard is assumed by Kaz USA, Inc., a Helen of Troy Company,
Health & Home, 400 Donald Lynch Blvd., Suite 300, Marlborough, MA 01752.
ASTM laboratory accuracy requirements in the display range of 37 to 39 °C
(98 to 102 °F) for IR thermometers is +/- 0.2 °C (+/- 0.4 °F), whereas for mercury-in-glass
and electronic thermometers, the requirement per ASTM Standards E667-86 and
E1112-86 is +/- 0.1 °C (+/- 0.2 °F).