14 Maintenance
Prior to working on the unit or in case of danger, disconnect it from the mains.
CAUTION
Risk of infection due to burst filters
Particles enter the compressed air network and can therefore enter the mouth of the patient.
Replace filters in accordance with the maintenance schedule.
14.1 Maintenance schedule
NOTICE
Risk of damage to the unit due to blocked filters
Continuous running due to reduced delivery. Damage to the unit due to burst filters.
Replace filters in accordance with the maintenance schedule.
Maintenance
interval
Maintenance work
At regular intervals
Empty the collector tray under the membrane drying unit (the interval may vary
depending on the ambient conditions and method of working; empty it daily if
the humidity is high).
Annually
Replace the air intake filter in the compressor unit – every six months if the
concentration of dust is high.
Replace the fine or sterile filter.
Replace the sintered filter.
Every 4 years
Replace the vibration dampers.
In accordance with
national law
Check the safety valve.
Carry out recurring safety inspections (e.g. pressure tank inspections, electrical
safety inspections) in accordance with applicable national laws.
Usage
36 5152100003L02 1910V001
EN

Other manuals for Durr Dental Primo

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Durr Dental Primo and is the answer not in the manual?

Durr Dental Primo Specifications

General IconGeneral
BrandDurr Dental
ModelPrimo
CategoryDental equipment
LanguageEnglish

Related product manuals