2008K
2
Preventive Maintenance Procedures
P/N 508138 Rev. J
Table 1. Alarm Tests.
ALARM TEST RESPONSE
Air Detector
1. Red visual alarm
2. Venous clamp closes
3. Blood pump stops
Blood Leak
1 Red visual alarm
2. Venous clamp closes
3. Blood pump stops
Arterial Pressure
1. Red visual alarm
2. Venous clamp closes
3. Blood pump stops
Venous Pressure
1. Red visual alarm
2. Venous clamp closes
3. Blood pump stops
TMP
1. Red visual alarm
2. Venous clamp closes
3. Blood pump stops
9-Volt Battery
Passes if battery voltage is greater than 7.0
volts under a load of 22.
Optical Detector
1. Lower venous alarm limit rises to 10mmHg
causing a venous pressure alarm
2. Venous clamp closes
3. Blood pump stops
Temperature
1. Red visual alarm
2. Yellow visual bypass
3. Machine in Bypass mode (no flow through the
dialysate flow indicator).
Conductivity
1. Red visual alarm
2. Yellow visual bypass
3. Machine in Bypass mode (no flow through the
dialysate flow indicator).

Other manuals for Fresenius Medical Care 2008K2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008K2 and is the answer not in the manual?

Fresenius Medical Care 2008K2 Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008K2
CategoryMedical Equipment
LanguageEnglish

Related product manuals