Page vi
2008K
2
Preventive Maintenance Procedures
P/N 508138 Rev. J
General Warnings
Warning! Never perform maintenance when a patient is
connected to the machine. If possible, remove the machine from
the treatment area when it is being serviced. Label the machine to
ensure it is not accidentally returned to clinical use before the
service work is completed. Always fully test the machine when
maintenance is completed. Confirm dialysate conductivity and pH
level before returning the machine to clinical use.
Shock Hazard: Refer servicing to qualified personnel.
The electrical source must be 120 volts, 60 Hz, single phase. The
outlet must be a three-conductor type with a hospital-grade
receptacle and a ground fault interrupter. Test the polarity and
ground integrity before installation and ensure it is maintained
thereafter. Failure to do so may result in electrical shock to the
operator or patient
Warning! Only Original Equipment Manufacturer (OEM)
Fresenius Medical Care parts should be used in the repair or
upgrade of the Fresenius Medical Care 2008K
2
Hemodialysis
System. Although, parts may look similar to parts in various
vendor catalogs or brick and mortar stores the 2008K
2
Hemodialysis System uses parts that have been specified and tested
in accordance to ANSI/AAMI/ISO guidelines. The use of non-
OEM parts will void your warranty and may cause patient harm.
Note: This document is written for the 2008K
2
Hemodialysis
Systems using software versions 5.28 or later.