Fresenius Medical Care 4008 HDF 3/11.97 (TM) 3-5
Select ADJ. UF-PUMP VOLUME by
pressing the ▲ and ▼ keys.
Press the Confirm key.
Press the Confirm key.
Select UF-PUMP 2 by pressing the
▲ and ▼ keys.
Press the Confirm key.
3.1 Calibrating the UF2 pump
Switch off hemodialysis machine.
Set service switch to ON (top).
Switch on hemodialysis machine.
Confirm
Select
CALIBRATION
☛
Confirm
Select
ADJ. UF-PUMP VOLUME
☛
Confirm
Select
UF-Pump 2
☛
Enter the number of strokes by pressing
the ▲ and ▼ keys.
Confirm
Select
pulse-amount = 60
☛
–40
0
–10
–20
–30
Single
Needle
Test
Prime
Override
☛
Press the Override key.
Text display for approx 3 sec.
Confirm
Select
ACKNOWLEDGED