Fresenius Medical Care 4008 HDF 3/11.97 (TM) 3-5
Select ADJ. UF-PUMP VOLUME by
pressing the and keys.
Press the Confirm key.
Press the Confirm key.
Select UF-PUMP 2 by pressing the
and keys.
Press the Confirm key.
3.1 Calibrating the UF2 pump
Switch off hemodialysis machine.
Set service switch to ON (top).
Switch on hemodialysis machine.
Confirm
Select
CALIBRATION
Confirm
Select
ADJ. UF-PUMP VOLUME
Confirm
Select
UF-Pump 2
Enter the number of strokes by pressing
the and keys.
Confirm
Select
pulse-amount = 60
40
0
10
20
30
Single
Needle
Test
Prime
Override
Press the Override key.
Text display for approx 3 sec.
Confirm
Select
ACKNOWLEDGED