Fresenius Medical Care 4008 HDF 3/11.97 (TM) 3-15
3.5 Calibrating the HDF blood pump
3.5.1 Pressure transducer
X348V_SN XHDF
XWAAGE
0
F
E
D
C
B
A
9
8
7
6
5
4
3
2
1
Switch the hemodialysis machine off.
Set HEX switch on the LP 754 to
position “F”.
Switch the hemodialysis machine on.
Start
Stop
Rate: ml/min
(Ø: mm)
HDFReset
Vol : l
Start
Stop
Rate: ml/min
(Ø: mm)
HDFReset
Vol : l
☛
☛
Display 000
Open SN pressure transducer to atmos-
phere.
Press the Start/Stop key.
Display 250
Apply a pressure of 250 mmHg to the SN
pressure transducer.
Press the Start/Stop key.