2062665-001 G 13
PCD-01: PCD-01 description
PCD-01 description
IHE Patient Care Device Domain > Device Enterprise Communication
(DEC) Profile > PCD-01 Transaction
IHE PCD-01 is an Integration Profile that simplifies and adds semantic meaning
to HL7 V2.6 messaging. The device conforms to PCD-01 as a Device Observation
Report (DOR) actor using:
ISO/IEEE 11073-10101 nomenclature in OBX-3 (Observation Identifier).
Virtual medical device containment tree in OBX-4 (Observation Sub-ID).
The originating medical device in OBX-18 (equipment instance identifier).
The value of this identifier is composed of two components. Component #1
is an EUI-64 identifier that starts with the GE Healthcare company code, plus
the EUI-64 separator, followed by the last 3 octects of the system's MAC
address in hex upper-case with no separators., plus a string that identifies
the monitoring equipment. For example, “080019FFFEAABBCCxxxxxxxx”.
Component #2 is “B40_GE”.
NOTE: The OBX-4 Observation Sub-ID follows PCD observation results
guidelines for relationships and elements in a hierarchical tree shown in
the following diagram.
Medical Device System (MDS)
Channel (Chan)
Metric
Numeric
Enumeration
Virtual Medical Device (VMD)
1
0..*
1
*
1
*
1
*

Table of Contents