2 Installation
WARNING:
DO NOT ATTEMPT TO CHARGE NON-RECHARGEABLE
BATTERIES. TO PREVENT ACCIDENTAL CHARGING OF NON-
RECHARGEABLE BATTERIES THE CHARGE LINK LK2 MUST BE
IN THE DISABLED POSITION (
).
C
AUTION:
D
O NOT LEAVE DISCHARGED BATTERIES IN THE INSTRUMENT.
O
LD BATTERIES CAN LEAK AND CAUSE CORROSION.
W
HEN FITTING THE BATTERIES MAKE SURE THE ELECTRICAL CONTACTS
ARE CLEAN AND OBSERVE THE CORRECT POLARITY.
Note
: For further information, refer to the safety page at
the front of this manual and to the battery
manufacturer.
2.1 Battery (Figure 2-1)
Three batteries are fitted in the battery compartment.
For non rechargeable batteries make sure the charging link
LK2 is fitted in the disabled position ( ).
For rechargeable batteries make sure the charging link LK2
is fitted in the enabled position (
).
Note
: The charging link LK2 is located on the instrument
PCB. To access this link remove the top case
assembly.
To replace the batteries, unscrew the captive screw and slide
off the battery compartment cover. Make sure the polarity of
the new batteries is correct.
If the battery capacity is low the display shows:
The battery symbol (approximately 1 hour left).
The flashing message Battery Very Low for 15 to 20
minutes.
The message Batteries Dead! followed, after a short time,
Switching Off!! the instrument then automatically switches
off.