DPI 740 User Manual 45
K0200 Issue No. 1
6.4Command Set
All commands are of the format described in 6.2. In the
following command description the start and terminating
characters have been removed for clarity. Parameters
enclosed in [ ] are optional.
Input commands
Command: IC - Set-up input
Format: IC=<param type>
Description: Set-up input to measure parameter <param
type>, the index `i` specifies the measurand.
<param type> can be any of the following:
P pressure input
I current input
V voltage input
T temperature input
Note
:Only use pressure P for this instrument.
Query: The corresponding query command allows
the interrogation of the currently selected
parameter.
Example: IC?
The instrument replies IC=P
(input = pressure)
Command:IR - Get input reading
Format: IR?
Description:Request for a reading
Example: IR?
The instrument replies IR=<value>
where: <value> is the reading in the
selected units (e.g., IR=1017.95)
Query: This is a query only command.