EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ F Series

GE LOGIQ F Series User Manual

GE LOGIQ F Series
167 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #149 background imageLoading...
Page #149 background image
GE LOGIQ F SERIES
DIRECTION 5446617-100, REVISION 10 BASIC SERVICE MANUAL
Section 10-3 - Maintenance Task Schedule 10 - 3
Section 10-3 Maintenance Task Schedule (cont’d)
NOTE: PMs are not mandatory, the table above is for reference only.
NOTE: May require specialized equipment to complete.
Table 10-2 Customer Care Schedule
Service at Indicated Time Daily Weekly Monthly
Per Facilities
QA Program
Notes
Clean Probe Holders
Clean Air Filter
more frequently depending on
your environment
Inspect AC Mains Cable
Mobile Unit Check Weekly
Inspect Cables and Connectors
Clean Console
Inspect Wheels, Casters, brakes and
Swivel Locks
Mobile Unit Check Daily
Check Control Panel Movement
Mobile Unit Check Daily
Console Leakage Current Checks
also after corrective
maintenance
Peripheral Leakage Current Checks
also after corrective
maintenance
Surface Probe Leakage Current Checks
also after corrective
maintenance
Endocavity Probe Leakage
Current Checks
also after corrective
maintenance
Transesphongeal Probe Leakage
Current Checks
also after corrective
maintenance
Surgical Probe Leakage
Current Checks
also after corrective
maintenance
Measurement Accuracy Checks
also after corrective
maintenance
Functional Checks
also after corrective
maintenance