GE LOGIQ V5/LOGIQ V3
D
IRECTION 5496012-100, REVISION 3BASIC SERVICE MANUAL
2 - 8 Section 2-2 - General Console Requirements
2-3-3 Minimal floor plan suggestion
CSI 8x10
Figure 2-1. Minimal floor plan, 2.5 m x 3 m (8 by 10 foot)
Scale:
Each square equals one square foot (app. 31 x 31 cm)
1. Door – at least 762 mm (30 inches)
2. Film Viewer
3. Counter Top, Sink with hot and cold
water and Supplies Storage
4. Linen Supply
5. Probes/Supplies
6. Examination Table – 1930 x 610 mm
(76 x 24 inches)
7. Footswitch
8. Stool
9. Ultrasound system
10. External Peripherals
11. Dedicated Power Outlet - Circuit
Breaker protected and easily
accessible
12. Network Interface
13. 457 mm (18 inches) distance of Ultrasound system
from wall or objects
14. GE Cabinet for Software and Manuals
4
1
3
11
6
13
8
10
5
7
9
12
14
2

Table of Contents